PRODUCTS

Technology for Future

Product

SB-4ClO₂-1 (ClO₂)

Electrochemical Chlorine Dioxide (ClO₂) Sensor (0–1 ppm)

The SB-4ClO₂-1 is an electrochemical gas sensor designed for precise detection of chlorine dioxide (ClO₂) within the 0 to 1 ppm range. It is suitable for both portable and fixed gas detection devices, offering linear output characteristics and excellent stability. With a fast T90 response time of less than 30 seconds, the sensor enables reliable ClO₂ monitoring across a wide range of industrial applications. It operates reliably between -20℃ and +50℃ and provides an expected operating life of more than 24 months in air.

KEY FEATURES
    • Measurement Range: 0–1 ppm
    • Output Signal: -800 ± 300 nA / ppm
    • Response Time: T90 < 30 sec
    • Resolution: ~0.05 ppm (electronics dependent)
    • Repeatability: ≤ ±5% of signal
    • Linearity: Linear
    • Operating Life: ≥ 24 months in air
    • Operating Temperature Range: -20℃ to +50℃
    • Operating Humidity Range: 15–90% RH
Product Inquiry

SB-4ClO₂-1 (ClO₂)

Electrochemical Chlorine Dioxide (ClO₂) Sensor (0–1 ppm)

Product Description

The SB-4ClO₂-1 is an electrochemical gas sensor designed for precise detection of chlorine dioxide (ClO₂) within the range of 0 to 1 ppm. It is suitable for both portable and fixed gas detection devices, offering linear output characteristics and excellent stability. With a fast T90 response time of less than 30 seconds, the sensor ensures reliable ClO₂ detection across various industrial applications. It operates stably over a wide temperature range from -20℃ to +50℃ and provides an expected operating life of more than 24 months in ambient air.


Key Features

- Measurement range: 0–1 ppm
- Output signal: -800 ± 300 nA / ppm
- Fast response time: T90 < 30 sec
- Resolution: approx. 0.05 ppm (electronics dependent)
- Repeatability: ≤ ±5% of signal
- Linearity: Linear
- Operating life: ≥ 24 months in air
- Operating temperature range: -20℃ to +50℃
- Operating humidity range: 15% to 90% RH


Product Specifications
Performance Characteristics
Output Signal-800 ± 300 nA / ppm
Typical Baseline Range (pure air, @ 20℃)≤ ±0.03 ppm ClO₂ equivalent
Baseline Shift (-20℃ ~ +50℃)≤ ±0.1 ppm typical
Response Time (T90)< 30 seconds
Measurement Range0–1 ppm
Maximum Overload10 ppm
LinearityLinear
Repeatability< ±5% of signal
Recommended Load Resistor10 ohms
Resolution (Electronics Dependent)< 0.05 ppm typical
Bias VoltageNot required
Environmental
Temperature Range Continuous-20℃ to +50℃
Pressure Range800 to 1200 mbar
Operating Humidity Range15% to 90% RH
Lifetime
Long Term Output Drift< 5% per annum
Recommended Storage Temp0℃ to 20℃
Expected Operating Life> 24 months in air
Standard Warranty12 months from date of dispatch
Intrinsic Safety Data
Maximum at 10 ppm15 µA
Maximum o/c Voltage< 1.0 V
Maximum s/c Current< 0.1 A
Sensor Specifications
Accessories
Download
PDF
Datasheet
Download